This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation and updates to h3 and h6.Additionally, the following corrections due to errors in the initial report: h5 (labeled for single use?) "yes" selected and h8 (usage of device) "initial use of device" selected.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Three attempts were performed to obtain additional information, but no response was received from the customer.Based on the results of the investigation, the non-insulated part falling off may have occurred due to the following: 1.The tissue pad experienced uneven wear due to the seal & cut mode activation, either by closing the grasping section without actually grasping any tissue (including after tissue resection), or by grasping and twisting the tissue.2.The uneven wear of the tissue pad led to the non-insulated area coming into contact with the distal end of the probe.3.The output was activated in seal & cut mode while the non-insulated area was in contact with the probe, resulting in an error.It also left a contact mark.The tb-0009of instruction manual provides the following for the prevention method of the associated with the event: warning ¿ do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.Warning ¿ when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.Olympus will continue to monitor field performance for this device.
|