|
Catalog Number EL5ML |
Device Problems
Failure to Form Staple (2579); Failure to Fire (2610)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Date sent: 3/28/2024 b3: unknown, assumed first day of month that complaint was reported d4: batch # x7046v a manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.Additional information was requested and the following was obtained: "no patient consequences." "how did device not work? impossible to release clips normally did device jammed (not fire clips)? yes did device not feed clips? the loaders were apparently well loaded did device drop or eject clips? yes, but after 5 attempts, the clips wouldn't hold.The clips were not securely closed did device sideways feed clips? no did device fire malformed clips? yes did device fire scissored clips? but not with scissors" attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
It was reported that during an unknown procedure, the device would not clip.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 5/21/2024.D4: batch # x7046v.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was returned with no apparent damage.In an attempt to replicate the reported event, the device was tested for functionality.Upon testing, the device was fired, and the clips did not advance into the jaw.The device was noted to have the tip of the advancer bent.The instrument was disassembled, upon disassembly the advancer was confirmed to be bent and eleven(11) clips were found inside clip track.However, it is known from the history of the device that bent advancer condition may lead to malformed clips.The event reported was confirmed and it is related to improper use of the device.Please note that prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.Please reference the instruction for use for more information.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
|
|
Search Alerts/Recalls
|
|
|