This report is 1 of 3 linked to mfg report numbers: 3004608878-2024-00040, 3004608878-2024-00041.A facility reported that there was an issue (movement) with the mayfield modified skull clamp (a1059) during craniotomy surgery.The patient was already under anesthesia when the dysfunction was observed; however, there was no consequences/impact for the patient.The surgery was finalized using a replacement unit with approximately 5 to 10 minutes increased surgery time (time to find another skull clamp).The only biographic information received is that the patient was an adult.
|
The mayfield skull clamp (a1059) was returned for evaluation: failure analysis - the inspection of the skull clamp shows it has rotational movement in its swivel lock assembly and a residue buildup was present.For the adjustment of the lock assembly, new components were required to replace worn internal parts.To resolve all issues, the internal parts of the unit were replaced to adjust it according to manufacturer specifications and to clean the skull clamp's swivel lock assembly.After completing the reassembly, including the adjustment, the skull clamp was subjected to a successful function test.Root cause - the complaint is confirmed via inspection of the unit.Probable root cause is routine use and wear.The unit is beyond integra¿s 7 years recommended life cycle and required replacement of worn internal components and cleaning of residue buildup.Additionally, improper, or suboptimal placement of the skull clamp can contribute to slippage/movement of the patient.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|