• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71953-01
Device Problem Premature Discharge of Battery (1057)
Patient Problems Hypoglycemia (1912); Loss of consciousness (2418); Convulsion/Seizure (4406)
Event Date 03/18/2024
Event Type  Injury  
Manufacturer Narrative
The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the freestyle libre reader were reviewed and the dhrs showed the freestyle libre reader passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A battery/power issue was reported with the abbott diabetes care (adc) device.The customer was unable to obtain readings due to a fast-draining battery and as a result, the customer became hyperglycemic with symptoms of seizure and a loss of consciousness.The customer had contact with a healthcare professional (hcp) and was provided electrolytes and glucose for treatment.No further treatment was required.There was no report of death or permanent impairment associated with this event.
 
Event Description
A battery/power issue was reported with the abbott diabetes care (adc) device.The customer was unable to obtain readings due to a fast-draining battery and as a result, the customer became hyperglycemic with symptoms of seizure and a loss of consciousness.The customer had contact with a healthcare professional (hcp) and was provided electrolytes and glucose for treatment.No further treatment was required.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre reader were reviewed and the dhrs showed the libre reader passed all tests prior to release.In addition, the review determined that the correct cable and adapter were part of the reader kit pack and there was no indication that the product did not meet specifications.If the product is returned, a physical investigation will be performed and a follow-up report submitted.This serves as a correction report.Section h11 (additional mfg narrative) was incorrectly submitted in the initial report.Correction has been made.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key18995113
MDR Text Key338830963
Report Number2954323-2024-10228
Device Sequence Number1
Product Code QLG
UDI-Device Identifier05021791002481
UDI-Public(01)05021791002481(10)06D296F(91)71965-01
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71953-01
Was Device Available for Evaluation? No
Date Manufacturer Received03/18/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-