Catalog Number 21029371001
Device Problems Mechanical Problem (1384); Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2024
Event Type  malfunction  
Manufacturer Narrative
The serial number of the integra 400 plus analyzer is (b)(6).The field service engineer determined there was a potential issue with the mixing tower seals.The mixing tower seals were replaced and ise tubing was adjusted.An ise performance check, calibration, control, and precision studies were performed.The investigation is ongoing.
 
Event Description
The initial reporter stated they received discrepant results for one patient sample tested with the sodium electrode on a cobas integra 400 plus analyzer.The customer noted they have been receiving frequent errors on the instrument after tubing had been replaced on the analyzer the week before.The sample initially resulted in a na value of 124 mmol/l.The doctor questioned the value, so the sample was repeated.The repeat na value was 135 mmol/l.The repeat value was deemed correct.
 
Manufacturer Narrative
Medwatch fields d1-d4, g1, and g4 were updated.The investigation determined the service actions resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SODIUM ELECTRODE
Type of Device
ELECTRODE, ION SPECIFIC, SODIUM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE INSTRUMENT CENTER AG TEGIMENTA
forrenstrasse
rotkreuz 6343
WZ   6343
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250-0457
MDR Report Key18995945
MDR Text Key338887947
Report Number1823260-2024-00931
Device Sequence Number1
Product Code JGS
UDI-Device Identifier04015630032457
UDI-Public04015630032457
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K951595
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/08/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number21029371001
Device Lot Number215303547
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/12/2024
Initial Date FDA Received03/28/2024
Supplement Dates Manufacturer Received04/23/2024
Supplement Dates FDA Received05/08/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ACETAMINOPHEN (TYLENOL EXTRA STRENGTH EZ TABLET); APIXABAN TABLET; FENTANYL TRANSDERMAL FILM; LEVOTHYROXINE TABLET; LORATADINE TABLET; LORAZEPAM TABLET; NALOXONE NASAL SPRAY; ODANSETRON TABLET; OMEPRAZOLE CAPSULE; OXYCODONE TABLET; PROCHLORPERAZINE TABLET; TRIAMCINOLONE TOPICAL CREAM
Patient Age53 YR
Patient SexMale
-
-