MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE INJECTOR; INJECTOR AND SHEATHSET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE INJECTOR; INJECTOR AND SHEATHSET Back to Search Results
Model Number NM-401L-0423
Device Problem Failure to Discharge (1169)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2024
Event Type  malfunction  
Manufacturer Narrative
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the olympus injection needle experienced an injection failure during multiple cases.Reportedly, the nurse had primed the needle prior to use without issue.However, after inserting the needle into the scope, it deployed, but no solution was injected due to a blockage.This event was noted during a diagnostic bronchoscopy procedure.According to the initial reporter, the procedure was completed using the subject devices with no reports of patient harm.
 
Manufacturer Narrative
This supplemental report is being submitted to provide a correction, the results of the device evaluation, and the results of the legal manufacturer's final investigation.Correction added to field: d4.Additional information added to fields: h2, h3, h4, and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.Based on the results of the investigation, it is likely that the friction between the outer tube and the needle increased by the following factors: the needle extended/retracted while the tube was coiled in inspection of operation.The slider was abruptly pushed.Angle of the distal end of the endoscope: a bending force might have been applied to the tube when the device was inserted into the endoscope, removed from the sterile package or during pre-inspection.This might have caused the tube to buckle; however, a root cause could not be determined.The event can be detected/prevented by following the instructions for use which state: straighten out the instrument before inspecting it.The instrument can be damaged if it is coiled while the handle is operated.Operate the slider slowly, otherwise the tube could buckle.Do not coil the insertion portion with a diameter of less than 15 cm.This could damage the insertion portion.When inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.Insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.Stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE INJECTOR
Type of Device
INJECTOR AND SHEATHSET
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18996069
MDR Text Key338984749
Report Number9614641-2024-00782
Device Sequence Number1
Product Code FBK
UDI-Device Identifier04953170260018
UDI-Public04953170260018
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K902736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNM-401L-0423
Device Lot Number23K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/20/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-