Model Number G151
Device Problem High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2024
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited high out of range pace impedance measurements on the right ventricular (rv) lead.The involved lead is a non boston scientific product.Boston scientific technical services (ts) recommended to continue to monitor.No adverse patient effects were reported.At this time, this device system remains in service.
 
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited high out of range pace impedance measurements on the right ventricular (rv) lead.The involved lead is a non boston scientific product.Boston scientific technical services (ts) recommended to continue to monitor.No adverse patient effects were reported.At this time, this device system remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18997932
MDR Text Key338913330
Report Number2124215-2024-18672
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534638
UDI-Public00802526534638
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/05/2019
Device Model NumberG151
Device Catalogue NumberG151
Device Lot Number122027
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/28/2024
Supplement Dates Manufacturer Received04/18/2024
Supplement Dates FDA Received04/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/30/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexMale
-
-