• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER TALISMAN PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER TALISMAN PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Catalog Number 9-PFO-3025
Device Problem Difficult to Fold, Unfold or Collapse (1254)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2024
Event Type  malfunction  
Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6) 2024, a 30-25mm amplatzer talisman pfo occluder was chosen for implant to close a patent foramen ovale (pfo) using a 9f amplatzer talisman delivery sheath.The patient had long tunnels greater than 13mm, and the defect was measured as 9.8mm with a sizing balloon.The occluder was deployed, but the right atrial disc was swollen like a mushroom during deployment.Pushing with the sheath did not improve the deformation.There was no interaction with cardiac structures during deployment, and there was no angulation or kink noticed in the delivery system.The device was removed and replaced with a 25-18mm amplatzer talisman pfo occluder.The procedure was completed successfully.There were no adverse patient effects.No health impact has been reported.
 
Manufacturer Narrative
An event of a device deforming during implantation was reported.A returned device assessment could not be performed as the device was not returned for analysis.Use of the incorrect size delivery system and anatomical interference are potential causes of the reported event.Information from the field indicated that the correct size delivery sheath was used but that there was interaction with cardiac structures during deployment of the second device.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER TALISMAN PFO OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18998421
MDR Text Key338950573
Report Number2135147-2024-01381
Device Sequence Number1
Product Code MLV
UDI-Device Identifier05415067033321
UDI-Public(01)05415067033321(17)260430(10)9026653
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P120021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9-PFO-3025
Device Lot Number9026653
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/28/2024
Supplement Dates Manufacturer Received04/09/2024
Supplement Dates FDA Received04/15/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/03/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
AMPLATZER TALISMAN DELIVERY SHEATH, 9-TDS-09F45-80
-
-