Catalog Number 1500275-33
Device Problems Failure to Advance (2524); Malposition of Device (2616); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2024
Event Type  malfunction  
Event Description
It was reported that the procedure was to treat a heavily calcified, moderately tortuous right coronary artery that is 80% stenosed.The vessel was prepared with orbital atherectomy and pre-dilatated with non-compliant 2.25mm and 2.5mm balloons.Buddy wire techinque was used with two non-abbott guide wires and a 2.75x33mm unspecified xience was attempted to be implanted at the distal portion of the lesion; however, was not able to advance all the way distally.The buddy wire was removed and the stent deployed from mid to proximal just after the first curve of the artery.Post dilatation was performed with a 3.0x12mm non-compliant balloon.A micro-catheter was used to deliver a 2.75x15mm xience sierra distally and post dilate with a non-compliant balloon.Finally, a 3.5x23mm xience sierra stent delivery system was attempted to be advanced; however, failed to cross due interaction with the 2.75x33mm xience stent that was previously implanted.Despite using the buddy wire, micro-catheter, further dilatation of the vessel, change of the sheath, the stent still failed to cross.Therefore, a non-abbott was used to successfully complete the procedure.There was no adverse patient effects and no clinically significant delay.No additional information was provided.
 
Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Manufacturer Narrative
The device was not returned for analysis.The lot history record and exception reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18998919
MDR Text Key338879737
Report Number2024168-2024-03871
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08717648218422
UDI-Public(01)08717648218422(17)260130(10)3021041
Combination Product (y/n)Y
Reporter Country CodeAS
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1500275-33
Device Lot Number3021041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/28/2024
Supplement Dates Manufacturer Received04/04/2024
Supplement Dates FDA Received04/22/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
3.5X23MM XIENCE SIERRA
-
-