Model Number CDVRA500Q
Device Problems Failure to Interrogate (1332); Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Event Description
It was reported that the implantable cardioverter defibrillator (icd) had gone into backup operation during or after a magnetic resonance imaging.No intervention was performed.Patient condition was stable.
 
Event Description
New information received notes that the implantable cardioverter defibrillator (icd) went into backup operation as magnetic resonance imaging setting was not enabled prior to imaging.It was also found that the icd was unable to be interrogated.A reset was performed, and the icd was restored.Patient condition was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19001150
MDR Text Key338884714
Report Number2017865-2024-37166
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067031990
UDI-Public05415067031990
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberCDVRA500Q
Device Lot NumberA000143852
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/21/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA MRI
Patient Age44 YR
Patient SexFemale
Patient Weight59 KG
-
-