• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COMPANION MEDICAL INC INPEN MMT-105NNBLNA NOVO NORDISK BLUE; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COMPANION MEDICAL INC INPEN MMT-105NNBLNA NOVO NORDISK BLUE; SYRINGE, PISTON Back to Search Results
Model Number MMT-105NNBLNA
Device Problems Computer Software Problem (1112); Material Too Rigid or Stiff (1544)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2024
Event Type  malfunction  
Manufacturer Narrative
Inpen paired with commercial mobile app with small dose.Received inpen [ 1/4 ] of travel.Performed encoder bond investigation and found that the encoder pattern wheel tabs rotating and traveling off the keyed slots of dose nut guides.Encoder pattern wheel should never rotate.This causes an unexpected travel of the encoder pattern wheel creating resistance to dial and dispensing.Also, plastic shavings contamination builds up found caused by encoder wheel tabs rubbing at the walls of the dose nut.Abrasions down the length of the dose nut.Unable to perform baseline/wireless functionality and displacement dose accuracy test.Performed leadscrew reset torque test.Inpen passed and is within specification (ccw: 3.70 ozf-in).Inpen failed front cap investigation.Inpen front shell does not fit securely onto cartridge holder due to small snap arm being cracked / broken.] in conclusion: it was determined from destructive analysis that the difficult to dial/dose was caused by a pattern wheel misalignment due to an encoder base bond failure.This can affect insulin delivery.Therefore, the customer concern of dose dial being difficult to dial/dose was confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported to medtronic minimed that the customer experienced inpen difficult to dial/dose, dose log/report data inaccuracy.The customer reported no adverse event.The event involved product(s) mmt-105nnblna.Troubleshooting was performed and observed that the customer reported dose knob/dial difficult to turn and a inpen replacement was initiated.Customer reported dose knob/dial difficult to turn.No harm requiring medical intervention was reported.Mmt-105nnblna was requested and customer response was the device will be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNBLNA NOVO NORDISK BLUE
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
lusine boyadzhyan
1223 world trade drive, 100
san diego, CA 92128
MDR Report Key19002797
MDR Text Key339261161
Report Number3012822846-2024-00391
Device Sequence Number1
Product Code FMF
UDI-Device Identifier000010862088000344
UDI-Public(01)000010862088000344(17)230907
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201337
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/07/2023
Device Model NumberMMT-105NNBLNA
Device Catalogue NumberMMT-105NNBLNA
Device Lot NumberB0593
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2024
Initial Date Manufacturer Received 03/15/2024
Initial Date FDA Received03/29/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Age19 YR
Patient SexMale
Patient Weight91 KG
-
-