• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problem Low Readings (2460)
Patient Problems Headache (1880); Hyperglycemia (1905)
Event Date 03/13/2024
Event Type  Injury  
Manufacturer Narrative
The product has been requested for investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A low readings issue with the adc device was reported.Customer received 65mg/dl lower sensor scan results.Due to low sensor scan results, customer was given treatment of grape juice and 2 sugar and as a result and experienced symptoms of "feeling drunk" and a headache.Customer was unable to self-treat and was taken to the hospital where treatment of insulin(dose/type unspecified) was given by a healthcare provider.A reading of 200mg/dl was also taken on a capillary test with unspecified time in relation to the event.No further treatment was reported.There was no report of death or permanent impairment associated with this event.
 
Event Description
A low readings issue with the adc device was reported.Customer received 65mg/dl lower sensor scan results.Due to low sensor scan results, customer was given treatment of grape juice and 2 sugar and as a result and experienced symptoms of "feeling drunk" and a headache.Customer was unable to self-treat and was taken to the hospital where treatment of insulin(dose/type unspecified) was given by a healthcare provider.A reading of 200mg/dl was also taken on a capillary test with unspecified time in relation to the event.No further treatment was reported.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs (device history review) for the fs libre sensor and fs libre sensor kits were reviewed and the dhrs (device history review) showed the fs libre sensor and sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report will be submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key19005324
MDR Text Key338934665
Report Number2954323-2024-10509
Device Sequence Number1
Product Code QLG
UDI-Device Identifier05021791001316
UDI-Public(01)05021791001316(17)2025-05-31(10)KTP008293(91)72002-01
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71992-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/18/2024
Initial Date FDA Received03/29/2024
Supplement Dates Manufacturer Received04/03/2024
Supplement Dates FDA Received04/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-