• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 620; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 620; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2024
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The flow sensors were replaced to resolve the issue.Block a: patient information could not be obtained due to country privacy laws.Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.
 
Event Description
It was reported that there was a malfunction resulting in a loss of mechanical ventilation.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 620
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH  214028
Manufacturer (Section G)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH   214028
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key19006108
MDR Text Key338937020
Report Number9710602-2024-00433
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier00840682103985
UDI-Public01008406821039857
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/04/2024
Initial Date FDA Received03/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/17/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-