• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problem Low Readings (2460)
Patient Problem Insufficient Information (4580)
Event Date 03/18/2024
Event Type  Injury  
Manufacturer Narrative
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The date of event is unknown.The date entered in section b3 is the date abbott diabetes care became aware of the event.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A low reading issue was reported with the abbott diabetes care (adc) device via social media ('facebook').The customer reported receiving unspecified scan results perceived to be low when compared to unspecified readings from a competitor brand meter.As a result, the customer reported they ¿ended up in the hospital¿ where they receive medical treatment from someone other than themselves.No further information was provided by the customer.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
This serves as a correction report.Section(s) updated as follows: g3, h2, h3, h6 and h11.H11: the most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned and a valid serial number has not been provided.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Clinical data was reviewed and confirmed that freestyle libre sensors continue to be safe, effective, and perform as intended in the field.Stability data for libre sensors was reviewed and showed no anomalies or non-conformances that could lead to the complaint.A tripped trend review was conducted for the reported complaint and freestyle libre sensors, no trends were identified that would indicate any product related issues.
 
Event Description
A low reading issue was reported with the abbott diabetes care (adc) device via social media ('facebook').The customer reported receiving unspecified scan results perceived to be low when compared to unspecified readings from a competitor brand meter.As a result, the customer reported they ¿ended up in the hospital¿ where they receive medical treatment from someone other than themselves.No further information was provided by the customer.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key19007033
MDR Text Key338978800
Report Number2954323-2024-10479
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71992-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/18/2024
Initial Date FDA Received03/29/2024
Supplement Dates Manufacturer Received03/29/2024
Supplement Dates FDA Received04/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-