Model Number 133650 |
Device Problem
Mechanics Altered (2984)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/04/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
According to the reporter, during n harvest of the great saphenous vein as coronary artery bypass grafting (cabg), open procedure, the device was misfiring and unable to be used.The surgeon was able to squeeze the handle.Individual self-loading liga applicators used in place.There was no patient injury.
|
|
Manufacturer Narrative
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the clips were not loading properly.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the device contains a ratchet mechanism to ensure that a clip is not allowed to release until a full handle squeeze is applied.As the handles are released, a new clip is automatically loaded into the clip applier jaw.After the last clip is used, the device is designed to lockout which prevents the handles from being closed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|