• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL EPIC¿ VALVE (MITRAL); HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL EPIC¿ VALVE (MITRAL); HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Catalog Number E100-27M
Device Problems Material Split, Cut or Torn (4008); Central Regurgitation (4068)
Patient Problem Mitral Valve Insufficiency/ Regurgitation (4451)
Event Date 03/01/2024
Event Type  Injury  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6) 2017, a 27mm epic mitral valve was successfully implanted.In (b)(6) 2023, the patient was presented with severe mitral regurgitation.On (b)(6) 2024, additional procedure was performed to explant the epic valve.A tear and prolapsed leaflets were observed upon open visual inspection.A replacement 27mm non-abbott device was implanted.The patient is reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC¿ VALVE (MITRAL)
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL LTDA. REG#3001883144
1301rua profvieira demendonça
bairro engenho nogueira 31.31 0-26
BR   31.310-260
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key19008326
MDR Text Key338980833
Report Number2135147-2024-01396
Device Sequence Number1
Product Code LWR
UDI-Device Identifier05414734027526
UDI-Public(01)05414734027526(17)200522(10)BR00007628
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P040021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/22/2020
Device Catalogue NumberE100-27M
Device Lot NumberBR00007628
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/05/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/23/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-