The sensor replacement alert was alerted 29 days after insertion, and an rma was issued for a sensor replacement.The returned sensor was tested in-house, and it revealed a loss of chemical performance which caused a decline in the signal modulation.The system correctly disabled the sensor due to performance deviation, and the system's self-test functions were working normally.The root cause of the failure was due to the sensor's hydrogel oxidation.As part of resolution, the user was inserted with a new sensor.B4.Date of this report updated to 11 april 2024.D9 device available for evaluation? yes,device received on 11/30/2023.G.Date received by the manufacturer? 09 april 2024.H3.Device evaluated by manufacturer? yes.H6.Type of investigation updated to 10.H6.Investigation findings updated to 3231.H6.Investigation conclusions updated to 4307.
|