• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 101967-950
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/08/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted retrospectively as part of internal review.The manufacturer is currently performing investigation and the results will be provided in the supplemental report.
 
Event Description
Senseonics was made aware of an incident where the user received an early sensor replacement alert resulting in early sensor removal.The user received the alert on 06 june 2023, on day 26 of sensor life.No adverse events were associated with this complaint.
 
Manufacturer Narrative
The sensor replacement alert was alerted 29 days after insertion, and an rma was issued for a sensor replacement.The returned sensor was tested in-house, and it revealed a loss of chemical performance which caused a decline in the signal modulation.The system correctly disabled the sensor due to performance deviation, and the system's self-test functions were working normally.The root cause of the failure was due to the sensor's hydrogel oxidation.As part of resolution, the user was inserted with a new sensor.B4.Date of this report updated to 11 april 2024.D9 device available for evaluation? yes,device received on 11/30/2023.G.Date received by the manufacturer? 09 april 2024.H3.Device evaluated by manufacturer? yes.H6.Type of investigation updated to 10.H6.Investigation findings updated to 3231.H6.Investigation conclusions updated to 4307.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown 20876-7005
MDR Report Key19009296
MDR Text Key339125552
Report Number3009862700-2024-00458
Device Sequence Number1
Product Code QHJ
UDI-Device Identifier00817491023308
UDI-Public817491023308
Combination Product (y/n)Y
Reporter Country CodeSP
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/23/2023
Device Model Number101967-950
Device Catalogue NumberFG-4500-50-302
Device Lot NumberWP08965
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received04/09/2024
Date Device Manufactured07/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-