MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-4000
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Event Description
The hospital reported that during an endoscopic vessel harvesting procedure, vasoview hemopro 2 heater wire separated from the jaw of the hemopro 2.The device remained intact.The jaws were closed during insertion into the cannula.A new device was opened to complete the procedure after a minimal delay.There was no patient harm.
 
Manufacturer Narrative
Tw id (b)(4) the device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Manufacturer Narrative
Trackwise#: (b)(4).Updated sections: b3, g3, g6, h2, h3, h6, h10.The device was returned to the factory for evaluation on 03/20/2024.A photograph was provided by the account.A photographic inspection was conducted.Signs of clinical use and evidence of blood and tissue were observed on the jaws.The clear silicone insulation of both the hot and cold jaws was observed to be intact with no visual defects.The heater wire was observed to be flexed away from the base of the hot jaw and detached from the tip of the jaw.An investigation was conducted on 3/26/2024.A visual inspection was conducted.Signs of clinical use and evidence of blood and charred material were observed on the jaws.A microscopic inspection was conducted.The clear silicone insulation of both the hot and cold jaws was observed to be intact with no visual defects.The heater wire was observed to be flexed away from the base of the hot jaw and detached from the tip of the jaw.Based on the returned condition of the device and investigation results, the reported failure "material twisted/bent; wire" was confirmed.The lot # 3000370626 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key19011317
MDR Text Key339007301
Report Number2242352-2024-00307
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public00607567700406
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 05/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Device Lot Number3000370626
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/12/2024
Initial Date FDA Received04/01/2024
Supplement Dates Manufacturer Received05/15/2024
Supplement Dates FDA Received05/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/05/2024
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN.
-
-