Catalog Number 05942861001
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/13/2024
Event Type  malfunction  
Manufacturer Narrative
The serial number of the customer's accu-chek inform ii meter is (b)(6).The customer stated that the qc levels 1 and 2 passed on the date of event.The test strips were requested for investigation but have not been received at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, accu-chek inform strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.
 
Event Description
We received an allegation of a questionable glucose result for one patient tested with the accu-chek inform ii meter.The meter result at 8:40 am was 35 mg/dl.The patient was treated based on this result.The laboratory result at 8:53 am was 55 mg/dl.
 
Manufacturer Narrative
The meter was received for investigation.The meter was tested using retention test strips and controls.Investigation results: control range: level 1 - 30-60 mg/dl.Results: level 1 ¿ 44, 43, 44 mg/dl.Control range: level 2 261-353 mg/dl.Results: level 2 ¿ 301, 305, 307 mg/dl.The investigation results are within acceptable range.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCU-CHEK INFORM II TEST STRIPS
Type of Device
BLOOD GLUCOSE MONITORING TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key19014138
MDR Text Key339053030
Report Number1823260-2024-00983
Device Sequence Number1
Product Code LFR
UDI-Device Identifier00365702428102
UDI-Public00365702428102
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K121679
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05942861001
Device Lot Number670943
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/13/2024
Initial Date FDA Received04/01/2024
Supplement Dates Manufacturer Received04/03/2024
Supplement Dates FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age1 DA
Patient SexMale
Patient Weight4 KG
-
-