Catalog Number 0684-00-0604 |
Device Problems
Leak/Splash (1354); Inappropriate Waveform (2536)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/12/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that during intra-aortic balloon (iab) therapy the patient was transferred in from another facility for an operation.There was no particular alarms during the operation.During icu management, after 2-3 days of therapy, the assist ratio was changed from 1:1 to 1:2 and ten minutes later a leak alarm was generated and blood was seen.It was also noted that the waveform was abnormal.The iab was removed after four days of therapy.The iab was replaced and therapy was continued.After removal, air was pushed into the catheter and a popping sound came from the balloon and a leak was discovered.There was no patient harm or adverse event reported.
|
|
Manufacturer Narrative
|
Event site postal code: (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Reference complaint #(b)(4).
|
|
Manufacturer Narrative
|
Device evaluation: the product was returned with the membrane completely unfolded and blood on the interior and exterior of the catheter and between the catheter and the sheath.The returned sheath was over the catheter but it was not a maquet product.The extender tubing was also returned.A portion of the extender tubing was cut off and not returned.The inner lumen was found occluded with dried blood.The occlusion was unable to be cleared.An underwater leak test of the balloon, catheter, y-fitting, extracorporeal and extender tubing was performed and two leaks were detected on the membrane at the same location approximately 9.4cm from the rear seal and both leaks measuring 0.025cm in length.The inner lumen was found occluded with dried blood.The occlusion was unable to be cleared.The reported problem was most likely triggered by leaks which was found on the membrane.The penetrations appear to have been caused by a sharp object.The penetrations were located at the same location which is an indication the balloon was folded and most likely occurred during iab insertion.A lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint # (b)(4).
|
|
Event Description
|
N/a.
|
|
Search Alerts/Recalls
|