It was reported that during a pulse field ablation (pfa) procedure, when the faradrive steerable sheath and farawave catheter were in the left atrium (la), it was noticed by the physician that there was a slow leak at the handle of the sheath.The point of the leak was at the "join" of the handle and the side port.Both the sheath and catheter were on pressure bags only (no pump) and the pressure was approximately 250mm/hg on each pressure bag.The physician inspected the device and there were no structural abnormalities on visual inspection of the sheath nor luer involvement with the leak.The physician replaced the device, and the issue was resolved.The procedure was completed.No patient complications were reported.The device has been returned for analysis.
|