• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRELINK; DATA MANAGEMENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRELINK; DATA MANAGEMENT SYSTEM Back to Search Results
Model Number 71733-01
Device Problems Device Alarm System (1012); Application Program Problem (2880)
Patient Problems Hypoglycemia (1912); Loss of consciousness (2418); Convulsion/Seizure (4406)
Event Date 03/16/2024
Event Type  Injury  
Event Description
An alarm issue was reported with the adc device unknown phone with unknown operating system version.Customer experienced a signal loss and was unable to obtain readings and receive glucose alarms.As a result, customer was not alerted of changes in glucose level and experienced seizure and loss of consciousness and was unable to self-treat.The customer had contact with a healthcare professional (hcp) who administered gel (unspecified) and glucose serum) requiring treatment for diagnosis of hypoglycemia.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
Sensor: the most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.Software: the most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The customer reported signal loss.Abbott diabetes care (adc) attempted to replicate the reported issue and the reported configuration of (iphone 13 mini, ios 17.0.3 , 2.10.2.7677) (samsung galaxy s20 fe 5g, android 13 ,2.10.1.10406).The reported issue was unable to be replicated and the system functioned as intended.There were no issues identified with the freestyle librelink app during replication that would have led to the reported issue.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The device manufacturing date does not apply.The date entered in section h4 is the date abbott diabetes care became aware of the event.Section d.Suspected medical device and g4 - pma/510(k)# has been populated for the freestyle librelink ios application as this report concerns a spain customer.This is same/similar to us freestyle libre 2 ios application, model number 71926-01.All pertinent information available to abbott diabetes care has been submitted.
 
Manufacturer Narrative
The freestyle librelink complaint was investigated and determined that there were no issues with the freestyle librelink application that would have led to the complaint.The user reported missing low glucose alarms.The reported issue was investigated and attempted to replicate using similar configuration and the reported issue was unable to be replicated and the system functioned as intended.Sensor (b)(6) has been returned and investigated.The sensor plug was properly seated, and no physical damage was observed on the sensor patch.The sensor was found to be in state 5 (indicating normal termination).Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.The reported sensor was activated with a retained reader and linearity testing was performed, all results were within specification.Both high and low glucose alarms were successfully activated.No malfunction or product deficiency was identified.This also serves as a correction report.Section (additional mfg narrative) was incorrectly updated in the previous report.Correction has been made.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
An alarm issue was reported with the adc device unknown phone with unknown operating system version.Customer experienced a signal loss and was unable to obtain readings and receive glucose alarms.As a result, customer was not alerted of changes in glucose level and experienced seizure and loss of consciousness and was unable to self-treat.The customer had contact with a healthcare professional (hcp) who administered gel (unspecified) and glucose serum) requiring treatment for diagnosis of hypoglycemia.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRELINK
Type of Device
DATA MANAGEMENT SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key19018562
MDR Text Key339094348
Report Number2954323-2024-10792
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71733-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received04/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2024
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-