MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number DBP-150SOLID145
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Embolism/Embolus (4438)
Event Date 12/18/2023
Event Type  Injury  
Event Description
It was identified that during an atherectomy procedure using a diamondback 360 peripheral orbital atherectomy device (oad), a type a dissection occurred.This was identified during an angiographic image review following the procedure.The procedure was completed with no adverse events reported.Additional information received from core lab on 3/19/2024 indicated a distal embolization occurred and was related to orbital atherectomy.
 
Manufacturer Narrative
H6 investigation conclusion code 22: the diamondback 360® peripheral orbital atherectomy system instructions for use manual states that distal embolization is a potential adverse event that may occur and/or require intervention with use of the system.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.Csi id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Type of Device
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
aaron stevens
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key19020459
MDR Text Key339123081
Report Number3004742232-2024-00155
Device Sequence Number1
Product Code MCW
UDI-Device Identifier10850026568490
UDI-Public(01)10850026568490(10)240630(10)436702-1
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K190634
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberDBP-150SOLID145
Device Catalogue Number7-10057-12
Device Lot Number436702-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/19/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age83 YR
Patient SexMale
Patient Weight55 KG
Patient RaceAsian
-
-