• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; NO MATCH Back to Search Results
Model Number CDVRA500Q
Device Problem Failure to Interrogate (1332)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Event Description
It was reported that the implantable cardioverter defibrillator exhibited bluetooth telemetry loss.No intervention was performed.There were no patient consequences reported.
 
Event Description
New information received notes that the patient was seen in clinic.A bluetooth reset was performed, and the pairing was restored.No patient consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19021332
MDR Text Key339135745
Report Number2017865-2024-37469
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067031990
UDI-Public(01)05415067031990(10)S000088704(17)241130
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDVRA500Q
Device Lot NumberS000088704
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/01/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/29/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA MRI.
Patient Age51 YR
Patient SexMale
-
-