MAUDE Adverse Event Report: EXELINT INTERNATIONAL, CO. EXEL 10-12ML SYR 20GX1" L-L; SYRINGE WITH NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXELINT INTERNATIONAL, CO. EXEL 10-12ML SYR 20GX1" L-L; SYRINGE WITH NEEDLE Back to Search Results
Model Number 26254
Device Problem Contamination (1120)
Patient Problem Phlebitis (2004)
Event Date 10/09/2020
Event Type  Injury  
Manufacturer Narrative
We have checked the retained samples as soon as below complaint was received, and no abnormal was found.This can be found in cmp (b)(4).
 
Event Description
The following is a legacy complaint that was investigated and closed in 2020 but is being re-opened in qualio because the incident should have originally been a reportable event.Reference capa-34 for explanation.While using this brand of needle syringe combo- multiple patients experiencing phelbitis- inflammation of injection spot.Then noticed while drawing different injections, black particles from the rubber stopper in syringe floating in solution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXEL 10-12ML SYR 20GX1" L-L
Type of Device
SYRINGE WITH NEEDLE
Manufacturer (Section D)
EXELINT INTERNATIONAL, CO.
2500 santa fe avenue
redondo beach CA 90278
Manufacturer Contact
vu doan
2500 santa fe avenue
redondo beach, CA 90278
3106490707
MDR Report Key19023036
MDR Text Key339154260
Report Number1035907-2024-00008
Device Sequence Number1
Product Code FMF
UDI-Device Identifier20221262544
UDI-Public20221262544
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 03/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/14/2023
Device Model Number26254
Device Lot Number181114
Was Device Available for Evaluation? No
Date Manufacturer Received10/09/2020
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Patient Sequence Number1
Patient Outcome(s) Other;
-
-