Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2b: additional product code: hwc.D9: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
It was reported that during a procedure using our tn-a, we had an issue with the aiming arm (03.043.029).After inserting nail, we attached the aiming arm and we attempted to drill for the first medial to lateral locking screw.When doing so, the drill was not aligned with the hole and was drilling into the nail.After removing the trocar and taking off the aiming arm and then reconnecting the aiming arm and re-inserting the trocars, the same issue occurred.It was last used on (b)(6) 2024 by the surgeon.No adverse events were reported with the outcome of the procedure, it did delay the case and we ended up having to do perfect circles proximally in the nail.This report is for one (1) aim-arm radioluc this is report 1 of 1 for complaint (b)(4).
|