Lot Number 233630
Device Problem Device Handling Problem (3265)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2024
Event Type  Injury  
Manufacturer Narrative
Based on the review of the batch documentation, lenstec can confirm that all procedures in the manufacturing and packaging of the lens and the cartridge were conducted correctly.Additionally, we have not received any other complaints from these batches.Furthermore, based on the investigation, the damage to the haptic and the optic suggests that the incorrect technique was used to load and eject the lens from the cartridge.Lenstec therefore recommends users to consult the ifu for the correct procedure and technique for loading and ejecting the lens.
 
Event Description
Lenstec, inc.Received an email notification stating "upon insertion, haptic broke.Surgeon removed and replaced.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFTEC HD
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church, BB170 92,
BB  BB17092,
Manufacturer (Section G)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church,
BB  
Manufacturer Contact
jimmy chacko
1765 commerce ave. n.
st. petersburg, FL 33716
7275712272
MDR Report Key19028641
MDR Text Key339208059
Report Number9613160-2024-00019
Device Sequence Number1
Product Code HQL
UDI-Device Identifier00844369027359
UDI-Public00844369027359
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P090022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Lot Number233630
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2024
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/09/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial