The reported events were lead dislodgement, sensing noise, failure to capture and a helix mechanism issue.As received, a complete lead was returned in one piece with the helix retracted and clogged blood/tissue.The reported events of sensing noise, failure to capture were not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.The reported event of a helix mechanism issue was confirmed.X-ray examination of the lead found the inner coil was over torqued in the connector region consistent with the procedure damage.After cleaning and by applying toque directly to the inner coil, the helix could be extended and retracted.The full helix extension length was measured to be within specification.The cause of the reported event of a helix mechanism issue was isolated to blood/tissue in the helix region and the inner coil over torqued in the connector region consistent with procedure damage.
|