Catalog Number D134701 |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/07/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that a patient underwent a supraventricular tachycardia (svt) ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter and the catheter was not irrigating.No ablations were performed.The physician removed the catheter from the patient and noticed no irrigation fluid was flowing through the tip of the catheter.There was no error noted from the irrigation pump.The correct setting was selected on the device.The physician administered heparin to the patient.When the catheter was replaced, the issue resolved, and the case continued.No adverse patient consequence was reported.
|
|
Manufacturer Narrative
|
If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
|
Manufacturer Narrative
|
The device investigation has been completed which included performing a manufacturing record evaluation (mre).The manufacturing record evaluation was performed for the finished device number lot 31215239l and no internal action related to the complaint was found during the review.Since no device has been received for analysis, no product investigation can be performed, and the customer complaint cannot be confirmed.However, if the product is received at a later date, the investigation will be updated as applicable.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
|
Search Alerts/Recalls
|
|