This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, it was determined that the reported condition that the device did not work was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the device had moving parts that did not move smoothly-trigger.The assignable root cause was determined to be due to component failure from wear.Udi: (b)(4).
|
It was reported by singapore that during service and evaluation, it was determined that the battery reamer drill device had moving parts that did not move smoothly-trigger.It was further determined that the device failed pretest for check for sticky trigger.It was noted in the service order that the device did not work.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|