Model Number 40027
Device Problems Incomplete Coaptation (2507); Central Regurgitation (4068)
Patient Problem Aortic Valve Insufficiency/ Regurgitation (4450)
Event Date 03/12/2024
Event Type  Injury  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
Medtronic received information that during the implant of this 27mm aortic bioprosthetic valve, it was reported that there was poor coaptation of the valve leaflets.It was stated that severe aortic insufficiency (ai) was present for about 5-10 minutes post implant after having seated the valve in and then closed the incision.It was further stated that the patient was put on a pump and the issue resolved instantly with the leaflets returning back to normal and no central jet observed.Central jet appeared then resolved on its own.It was noted that the device had been inspected prior to use.It was mentioned that the patient was discharged and is doing well.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVALUS AORTIC TISSUE VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key19031926
MDR Text Key339235479
Report Number2025587-2024-02065
Device Sequence Number1
Product Code LWR
UDI-Device Identifier00643169983212
UDI-Public00643169983212
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P170006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/03/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number40027
Device Catalogue Number40027
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/27/2024
Date Device Manufactured08/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age68 YR
Patient SexMale
-
-