It was reported that the procedure was to treat a moderately calcified, mildly tortuous peroneal artery that is 80% stenosed.The 2.5x60mm armada 14 balloon catheter was not air aspirated outside the anatomy prior to use.Once at the lesion, the balloon would not hold pressure after inflation to nominal pressure and was noted to be deflating.However, balloon angioplasty was completed with the same armada and then the device was removed.There was no adverse patient effects and no clinically significant delay reported.No additional information was provided.
|