Model Number SBT-114-100
Device Problem Material Separation (1562)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 03/12/2024
Event Type  malfunction  
Manufacturer Narrative
Block d4 and h4: the complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Block h6: imdrf device code a0413 captures the reportable event of bite blox material separation.
 
Event Description
It was reported to boston scientific corporation that a bite blox was used during an upper endoscopy performed on (b)(6) 2024 for the treatment of esophageal dilation.During the procedure, the bite blocks broke when scope reached the stomach for visualization.The scope was removed from the patient's mouth and noticed that a piece of the bite block was inside the patient's mouth with traces of blood.The detached piece was retrieved by hand.The procedure was aborted due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOX
Type of Device
ENDOSCOPIC BITE BLOCK
Manufacturer (Section D)
JIANGSU JIANYU HEALTH MEDICAL CO LTD
zhixi industry zone
jintan area
changzhou city 21325 1
CH  213251
Manufacturer (Section G)
JIANGSU JIANYU HEALTH MEDICAL CO LT
no 88 of longxi avenue
zhulin town
changzhou city 21324 1
CH   213241
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key19034502
MDR Text Key339282911
Report Number3005099803-2024-01492
Device Sequence Number1
Product Code MNK
UDI-Device Identifier00816849010168
UDI-Public00816849010168
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSBT-114-100
Device Catalogue NumberSBT-114-100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/12/2024
Initial Date FDA Received04/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1