THORATEC CORPORATION HEARTMATE 3 LVAS IMPLANT KIT; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 106524US
Device Problems Decreased Pump Speed (1500); Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2024
Event Type  malfunction  
Event Description
It was reported that on (b)(6) 2024 the patient required a stent emergently placed to the outflow graft.Repeat computed tomography angiography (cta) done on (b)(6) 2024 showed patency of the stent.Despite this, patient had low flow alarms over the weekend with 0 speed and 0 flow.Log files were sent for review, and it was recommended to try rebooting the system monitor.The customer tried to reboot the monitor without resolution and switched out the monitor and utilized the heartmate touch.
 
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 3 LVAS IMPLANT KIT
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key19035813
MDR Text Key339287208
Report Number2916596-2024-01728
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00813024013297
UDI-Public(01)00813024013297(10)6708529(17)210801
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/01/2021
Device Model Number106524US
Device Lot Number6708529
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/12/2024
Initial Date FDA Received04/03/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/25/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
Patient Weight78 KG
-
-