Catalog Number 72290038
Device Problem Self-Activation or Keying (1557)
Patient Problem Burn(s) (1757)
Event Date 03/07/2024
Event Type  malfunction  
Event Description
It was reported that, during an arthroscopy, when the werewolf coblation 90 wand was used, it continued to heat up even though the operator had released the button.This resulted in an intra-articular burn.Surgery was completed with a back-up device instead, it is unknown if there was any surgical delay.No further complications were reported.
 
Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Manufacturer Narrative
The reported device was received for evaluation.A visual inspection revealed the returned instruments power connector and fluid evacuation lines have been cut from the device.Bio debris present.Product was out of the original packaging.No packaging returned.A functional evaluation of the device could not be performed due to the power connector and fluid evacuation lines being cut from the device.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Factors that can contribute to the reported event include saline/humidity entered the interior of the handle shorting the connection of the buttons.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEREWOLF FLOW 90 COBLATION WAND
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19038911
MDR Text Key339321080
Report Number3006524618-2024-00133
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00885556647462
UDI-Public00885556647462
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K183346
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/03/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72290038
Device Lot Number2134173
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/12/2024
Date Device Manufactured10/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-