COCHLEAR LTD NUCLEUS CI512 COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI512
Device Problem Circuit Failure (1089)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This report is submitted on april 04, 2024.
 
Event Description
Per the clinic, open circuits were noted on 9 electrodes and the patient also experienced subsequent loss of connection to the internal device.Reprogramming attempts were made; however, the issue could not be resolved.There are plans to explant the device and to reimplant the patient with a new device; however, this has not occurred as of the date of this report.The implanted device remains.Additional information has been requested but has not been made available as of the date of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI512 COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS  2109
Manufacturer (Section G)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS   2109
Manufacturer Contact
nurshahqifah latif
unit ug-1, vertical podium
no. 8 jalan kerinchi
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key19039084
MDR Text Key339318663
Report Number6000034-2024-01217
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502014564
UDI-Public(01)09321502014564(11)220603(17)240602
Combination Product (y/n)N
Reporter Country CodeTC
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/02/2024
Device Model NumberCI512
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/12/2024
Initial Date FDA Received04/04/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-