This report is being supplemented to provide additional information based on the legal manufacturer's device evaluation and investigation.Based on the results of the device evaluation, the reported event was not confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 months since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the reported event could not be identified.However, it¿s likely the cause is related to a probe damage error by way of an excessive load being applied to the probe.The possible causes of the overload on the probe include the following: 1.Pressing against a hard object.2.Grasping thick tissue.3.Twisting while grasping the tissue.The event can be prevented by following instructions for use which state: ·do not activate output while the grasping section is closed without contacting tissue or vessel or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.·when cutting or vessel sealing is performed, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ during the treatment, do not activate output while applying the probe tip to the tissue with a strong force, grasping thick tissue, or twisting the handle.Also, do not insert the handle while the handle is twisted with respect to the tissue, do not gras it, and do not activate the output.Otherwise, the probe tip and/or grasping section may be damaged, which may result in falling of the probe tip and/or tissue pad.Olympus will continue to monitor field performance for this device.
|