Catalog Number 4824462
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2024
Event Type  malfunction  
Manufacturer Narrative
Investigation: the run data file (rdf) was analyzed for this event.The signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was possibly the result of a wbc saturation of the lrs chamber which occurred toward the end of the procedure.The signals in the run data file indicate that the device had difficulty maintaining steady state operation during the procedure, so it cannot be ruled out that the needle line access may have shifted at around 65 minutes into the procedure, and could potentially have impacted the collection.Additionally, based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product may have been donor-related.Investigation is in process, a follow-up report will be provided.
 
Event Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.Donor unit #: (b)(6) there was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA PLATELET+SAMPLER, PLASMA, RBC SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key19043423
MDR Text Key340119416
Report Number1722028-2024-00119
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583824461
UDI-Public05020583824461
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK190332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4824462
Device Lot Number2308016142
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/12/2024
Initial Date FDA Received04/04/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-