• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/13/2024
Event Type  malfunction  
Event Description
During the training session, the autopulse platform (b)(6) displayed user advisory (ua) 45 (not at "home" position after power-on/restart) error message.The user was unable to clear the error.No patient involvement.
 
Manufacturer Narrative
Zoll has not received autopulse platform for investigation.A supplemental report will be filed when the product is returned, and investigation has been completed.
 
Manufacturer Narrative
The reported complaint that the autopulse platform (sn (b)(6)) displayed user advisory (ua) 45 (not at "home" position after power-on/restart) was confirmed during the archive data review and functional testing.The root cause of the (ua) 45 advisory message was that the driveshaft was not in the "home" position, most likely attributed to unintended user error.Upon visual inspection, no physical damage was noted.The archive data show the presence of a user advisory (ua) 45 error message around the customer's reported date that likely related to the drive shaft not being in the home position during the power on, thus confirming the reported complaint.A user advisory is normally a clearable error message, and it is designed into the autopulse platform to alert the operator that autopulse has detected one of several conditions.Per the autopulse® resuscitation system model 100 user guide, if the driveshaft is not at its home position when the autopulse is powered on, a user advisory (45) will occur.This user advisory will persist until the driveshaft is returned to its home position.To clear a user advisory (45), pull up on the lifeband until the chest bands are fully extended (thereby moving the driveshaft back to its home position), and then restart.Functional evaluation failed as the autopulse platform displayed a (ua) 45 advisory message upon powering up, confirming the reported complaint.The driveshaft was rotated to the "home" position to remedy the fault.Following the service, the autopulse platform passed the run-in test without any fault or error.The autopulse platform passed the final testing without any fault or error.Historical complaints were reviewed for service information related to the reported complaint, and there was no previous history of complaints reported for the autopulse platform with sn (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa sackrison
2000 ringwood ave.
san jose, CA 95131
MDR Report Key19047537
MDR Text Key339435795
Report Number3010617000-2024-00289
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/13/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-