Reportable upon analysis completed 21march2024.It was reported that this microcatheter had kinked.During imaging of a renal embolization procedure using a direxion microcatheter, it was found that the device was faulty during tracking though the anatomy and the hub had kinked.It was noted that the carrier tube was hydrated prior to the removal of the catheter; however, there was a little resistance removing the catheter from the carrier tube.A new device was used to complete the procedure, and no patient complications were reported.However, device analysis revealed a fracture and a device separation.
|