SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problems Mechanical Problem (1384); Overheating of Device (1437); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/16/2021
Event Type  malfunction  
Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.A functional evaluation revealed a blade stall error and overheating of the housing.Further evaluation revealed the drive fork was stiff when rotated.The root cause was associated with a mechanical component failure.Factors that could have contributed to overheating include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This could be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.Several corrective actions were taken to reduce the occurrence of this failure.No further actions are required at this time.
 
Event Description
It was reported that during set up, the mdu was not working.There was a back-up device available and no delay or patient injuries were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19049983
MDR Text Key339474042
Report Number1643264-2024-00229
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2024
Initial Date Manufacturer Received 04/02/2024
Initial Date FDA Received04/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-