The hospital reported that during an endoscopic vessel harvesting procedure, vasoviewhempro vh-3500 wire within the cautery jaws separated from the jaw.The tool was oriented tips up when inserted into the cannula as trained.Another vh-3500 was opened to successfully complete the procedure.There was no significant time delay, just a few minutes to open another kit.There was no patient injury.
|
Trackwise #: (b)(4).Updated sections: b-4, g-3, g-6, h-2, h-6, h-10.A lot history record review was completed for lots 3000369005, 3000369645, and 3000367542 the last 3 lots shipped to the account prior to the event/aware date.There were no ncmrs, rework, or deviations documented for the last 3 lots shipped to the account.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|