ABBOTT DIABETES CARE INC FREESTYLE LIBRELINK; DATA MANAGEMENT SYSTEM Back to Search Results
Model Number 71732-01
Device Problems Device Alarm System (1012); Application Program Problem (2880)
Patient Problem Hypoglycemia (1912)
Event Date 03/28/2024
Event Type  Injury  
Event Description
An alarm issue was reported with the adc device in use with a huawei p 20 phone with an android operating system version 12.Customer experienced a signal loss and was unable to obtain readings and receive glucose alarms.As a result, customer was not alerted of changes in glucose level and was treated with orange juice and sugar by a non-healthcare provider.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
Sensor: the most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.Application: freestyle librelink complaint was investigated and no issues with the librelink application in use with huawei p 20 phone with an android operating system version 12 was determined that would have led to the complaint.The user reported signal loss alarm with the freestyle librelink application.Attempted to replicate the issue using similar device configuration and was not able to reproduce the complaint.If the product is returned, a physical investigation will be performed and a follow-up report submitted.Section d.Suspected medical device and g4 - pma/510(k)# has been populated for the freestyle librelink [ios/android] application as this report concerns a [complaint country] customer.This is same/similar to us freestyle libre 2 ios/android application part number [71926-01/71857-01].The device manufacturing date does not apply.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Manufacturer Narrative
The user reported signal loss.The reported issue was investigated and attempted to replicate.The reported configuration was not compatible with the freestyle librelink app.The latest revision of the compatibility guide was available to the customer on the abbott diabetes care website.Sensor (b)(6) has been returned and investigated.The sensor plug was properly seated, and no physical damage was observed on the sensor patch.The sensor was found to be in sensor state 4 (indicating the sensor had a remaining life of 12 hours before ending).Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.The sensor was activated with a known good reader and the bluetooth connection was successful.A linearity test was performed while the reader was wrapped in aluminum foil (to simulate signal loss), and signal loss message was observed.Sensor was scanned with reader to re-establish bluetooth connection and then was placed at a distance from the sensor.Signal loss message was not displayed.Therefore, the issue is not confirmed.This also serves as a correction report.Section (additional mfg narrative) was incorrectly updated in the previous report.Correction has been made.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
An alarm issue was reported with the adc device in use with a huawei p 20 phone with an android 12 operating system with app version 2.11.1.10973.Customer experienced a signal loss and was unable to obtain readings and receive glucose alarms.As a result, customer was not alerted of changes in glucose level and was treated with orange juice and sugar by a non-healthcare provider.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRELINK
Type of Device
DATA MANAGEMENT SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key19051979
MDR Text Key339497152
Report Number2954323-2024-11661
Device Sequence Number1
Product Code PZE
UDI-Device Identifier00357599000042
UDI-Public00357599000042
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71732-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 03/28/2024
Initial Date FDA Received04/05/2024
Supplement Dates Manufacturer Received04/10/2024
Supplement Dates FDA Received04/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/22/2024
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-