• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR Back to Search Results
Model Number G126
Device Problems High impedance (1291); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Electric Shock (2554)
Event Date 03/14/2024
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited high out of range impedance measurements on the right ventricular (rv) lead.A non-invasive programmed stimulation (nips) test was performed and as a result, a code 1005 indicative of an open circuit condition was displayed.A revision procedure was scheduled.Further information was received that the physician opted to continue to monitor this patient.No additional adverse patient effects were reported.At this time, this lead remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19053756
MDR Text Key339551927
Report Number2124215-2024-20845
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589140
UDI-Public00802526589140
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/24/2023
Device Model NumberG126
Device Catalogue NumberG126
Device Lot Number106143
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age69 YR
Patient SexFemale
-
-