• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 SITE-RITE 8 ULTRASOUND SYSTEM 20MM PINPOINT GT PROBE (REMAN); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 SITE-RITE 8 ULTRASOUND SYSTEM 20MM PINPOINT GT PROBE (REMAN); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Catalog Number 9770602Y
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Manufacturer Narrative
H11: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.Found during evaluation at bd service facility: the probe has a very rainy image due to an internal failure of the probe.H3 other text : evaluation summary findings in h:11.
 
Event Description
Observation found during evaluation at bd service facility: the probe has a very rainy image due to an internal failure of the probe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND SYSTEM 20MM PINPOINT GT PROBE (REMAN)
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
shelly gilbert
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key19056174
MDR Text Key340234197
Report Number3006260740-2024-01506
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741138201
UDI-Public(01)00801741138201
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182281
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9770602Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-