MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS VASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS VASCULAR STENT GRAFT Back to Search Results
Catalog Number FVL10080
Device Problems Break (1069); Positioning Failure (1158); Fracture (1260); Misfire (2532)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2024
Event Type  malfunction  
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the fluency plus endovascular stent graft products that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft products are identified in d2 and g4.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.However, photo was provided for review.The investigation of the reported event is currently underway.H10: d4: (expiry date: 08/2026).H11: section a: through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a stent graft placement procedure via right femoral artery approach in supine position, the stent conveyor rod was allegedly broken.It was further reported that the stent was allegedly could not be released normally.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the fluency plus endovascular stent graft products that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft products are identified in d2 and g4.H10: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was returned for evaluation; the stent graft was found partially deployed and the outer sheath was fractured which leads to confirmed results.The condition of the returned sample depicts a device which failed during active usage as opposed to the reported information that a breakage happened when the package was opened.Also, a photo was provided which showed a fractured outer sheath of the device.Based on provided photos and the evaluation of the return sample, the investigation is closed with confirmed results for sheath fracture and partial deployment.A definite root cause for the reported event could not be determined.The intended use of the device in the femoral artery represent an off-label use.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.Regarding preparation of the device the instructions for use states that "prior to loading the vascular system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment.Continue until saline drips from the distal end of the delivery system".Regarding precautions, instructions for use states: "if excessive force is felt during stent graft deployment, do not force the delivery system.Remove the delivery system and replace with a new unit".Regarding accessories the instructions for use states: "a super stiff guide wire (0.035 in.) is advanced from a femoral artery puncture site.Use an introducer sheath for the implant procedure; the packaging pictograms indicate an introducer size of 9f and a 0.035" guidewire.Predilatation of the stenosis may be performed prior to stent graft deployment at the discretion of the physician".The fluency plus vascular stent graft is indicated for the treatment of atherosclerotic lesions in the iliac arteries.H10: d4 (expiry date: 08/2026), g3, h6 (device, method).H11: b5, h6 (patient, result, conclusion).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that prior to a stent graft placement procedure for left superficial femoral artery occlusion via the right femoral artery approach, the stent conveyor rod was allegedly found to be broken.It was further reported that the stent was allegedly could not be released normally.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS VASCULAR STENT GRAFT
Type of Device
VASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key19056537
MDR Text Key339890433
Report Number9681442-2024-00070
Device Sequence Number1
Product Code PFV
UDI-Device Identifier00801741145230
UDI-Public(01)00801741145230
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P130029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFVL10080
Device Lot NumberANHV0433
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received04/06/2024
Supplement Dates Manufacturer Received04/18/2024
Supplement Dates FDA Received04/30/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-