H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the fluency plus endovascular stent graft products that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft products are identified in d2 and g4.H10: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent graft delivery system was returned for evaluation; the stent graft was completely deployed and missing while the outer sheath was fracture which leads to confirmed results for sheath fracture.Based on reported information and evaluation of the returned sample, the investigation is closed with confirmed results for sheath fracture.A definite root cause for the reported event could not be determined.Labeling review: relevant labeling was reviewed.It was found that the instructions for use sufficiently address the potential risks.Regarding anatomy the instructions for use states "if excessive force is felt during stent graft deployment, do not force the delivery system.Remove the delivery system and replace with a new unit".Regarding preparation of the device the instructions for use states: the instructions for use states: "prior to loading the delivery system over a guide wire, both ports must be flushed with sterile saline to eliminate any air bubbles that may be trapped in the inner catheter lumen and/or the stent graft lumen.Flushing these lumens will also facilitate stent graft deployment.A super stiff guide wire (0.035 in.) is advanced from a femoral artery puncture site.Use an introducer sheath for the implant procedure".The packaging pictograms indicate an introducer size of 10f and a 0.035" guidewire.Predilatation of the stenosis may be performed prior to stent graft deployment at the discretion of the physician." h10: d4 (expiration date: 07/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|