Catalog Number UNKNOWN |
Device Problem
Infusion or Flow Problem (2964)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 03/14/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that unspecified bd infusion set was occluded the following information was received by the initial reporter with the following verbatim: the charge nurse in labor and delivery reported that once or twice she has been infusing pitocin as a primary infusion, y-sited into a primary infusion of lr.The lr pump module was infusing without alarm, however the pitocin pump module was alarming ¿patient side occlusion¿.We educated her on the proper way to make a y-site connection.Primary tubing.
|
|
Manufacturer Narrative
|
B3.The date received by manufacturer has been used for this field.H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
|
Event Description
|
No additional info.
|
|
Manufacturer Narrative
|
No product or photo was returned by the customer.The customer complaint of flow issues - fluid blockage could not be verified due to the product not being returned for failure investigation.A device history record review could not be performed because the lot number is unknown.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.
|
|