It was reported that the patient with the cardiac resynchronization therapy defibrillator (crt-d) device exhibited high out of range pacing impedance measurements, measuring at 2019 ohms on this left ventricular (lv) channel.The impedance measurements were later stable in the range of 400 ohms; however, the threshold value went up.Technical services (ts) reviewed and stated that there was little noise visible on the presenting electrogram (egm) but it was not oversensed.The noise was able to be reproduced in the clinic.This lv lead remains in service.No adverse patient effects were reported.
|