SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID
|
Back to Search Results |
|
Catalog Number HL-90 |
Device Problems
Excessive Heating (4030); Key or Button Unresponsive/not Working (4063)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event is unknown, no information has been provided to date.Device evaluation: one device was returned for investigation.Visual inspection noted the device was received in good condition.Functional testing found the over temperature alarm test button didn't trigger the alarm confirming the customer complaint.Root cause was attributed to the over temperature alarm wasn't at the proper setting.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.The service history review identified there was no indication that the complaint was related to a service of the device within the review period.The over temperature alarm was reset.
|
|
Event Description
|
It was reported that the over temperature button is not working.This was found when they received the device back and bench testing was completed.There was no patient involvement.During over temperature (ot) testing, the device did not go into ot alarm with the push of the button.Patient did not receive medical treatment.There was no patient or clinical injury.
|
|